Raymarine ST60 Specifications Page 41

  • Download
  • Add to my manuals
  • Print
  • Page
    / 62
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 40
29
Chapter 4: Calibration
4.1 Introduction
The ST60+ Tridata instrument is set up with factory-programmed default settings,
so in order to optimise the performance of the instrument on board a particular
vessel, the procedures in this Chapter must be carried out immediately after the
completion of installation and before the equipment is used for navigational
purposes.
Where practicable, the calibration procedures are presented diagrammatically to
show the sequence of button presses and the resulting displays. Adjustment
instructions are given as applicable.
4.2 User calibration
The User calibration procedures enable you to:
Set the required units for depth readings.
Set the offset for depth readings, i.e. determine whether depth readings are
from the keel of the vessel or from the water line.
Lock the shallow alarm.
Set the required units for speed readings.
Set the speed resolution.
Select the log distance units
Set the speed readings on the instrument to be a true indication of the actual
speed of the vessel.
Select temperature units.
Calibrate for correct temperature readings.
Set timer alarm buzzer on or off.
Separate routines are provided for the User calibration of the depth and speed
functions. To carry out either of these routines:
1. Power up the ST60+ Tridata instrument.
2. Hold down the
depth and speed buttons for approximately 2 seconds so
that the User calibration entry screen is displayed.
3. Carry out the User calibration procedures for
Depth
and
Speed
.
81263_2.book Page 29 Monday, December 5, 2005 9:08 AM
Page view 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 61 62

Comments to this Manuals

No comments